Optical Excellence in the Laboratory: The Master Guide for the 3nh YS6002 Benchtop Color and Haze Meter
In modern quality assurance of plastics, films, glass, and liquids, subtle nuances determine product approval. The 3nh YS6002 Benchtop Color and Haze Meter combines high-precision spectrophotometry with standard-compliant haze measurement according to international standards. This practical guide shows you how to unleash the full potential of this high-end instrument, eliminate measurement errors, and generate reproducible quality data for your documentation.
Product Overview & Technical Flexibility
| Parameter / Feature | YS6002 Model Specification | Practical Benefit for Users |
|---|---|---|
| Optical Geometry | Transmissive: d/0 (Diffuse illumination, 0° viewing) Reflective: d/8 (Model dependent) |
Accurate haze and transmission measurements according to CIE standards. Prevents stray light interference. |
| Light Source & Sensor | 360–780 nm Full Spectrum Combined LED & UV Excitation | Extremely durable (>3 million measurements), precisely simulating the D65 light source. |
| Measuring Apertures | Standard: Ø 25.4 mm (Optional micro-measurement down to Ø 4 mm) | High flexibility for changing sample geometries and ultra-small sample quantities. |
| Sensor Architecture | Concave Grating & Dual-Array 256-Pixel CMOS Image Sensor | Highest optical resolution and spectral separation within 10 nm intervals. |
Laboratory-Grade Performance
Want to take your color and haze measurement to the next level?
Explore all technical specifications and secure the 3nh YS6002 for your quality assurance workflow.
View the 3nh YS6002 in the Shop NowRecommended Basic Settings for Standard Applications
Note: The following parameters serve as application-specific reference values and must be aligned with your internal testing protocols.
- Standard Illuminant: D65 (for classic daylight evaluations) or Illuminant A / C for specific haze testing according to ASTM D1003.
- Standard Observer: 2° for strict color analysis or 10° for large-scale visual correspondence.
- Measurement Mode: Quality Control Mode (QC Mode) for direct comparison of standard and sample with target tolerances.
- Color Spaces: Activate CIE Lab, XYZ, and Yxy to capture both haze factors and absolute color coordinates simultaneously.
The Professional Workflow: 6 Steps to Error-Free Measurements
- Power On and Warm-up: Start the device and allow a warm-up phase of approximately 15–20 minutes. The integrated temperature sensor monitors system stability.
- Perform Dual-Calibration: Meticulously execute black and white calibration using the provided transmission and reflection standards. This eliminates thermal drifts.
- Prepare the Sample Chamber: Choose the appropriate sample holder (e.g., the micro-aperture holder for small quantities) and clean the chamber of any residue.
- Define the Standard: Place your reference sample (or select a digitally stored standard profile) and complete the reference measurement.
- Start Sample Measurement: Secure the test sample (film, glass plate, or cuvette) absolutely flat and vibration-free in the chamber. Trigger the measurement via the 7-inch touchscreen.
- Data Export & QC Check: Verify the data metrics (Haze, WI, YI, Gardner, etc.) on the display and transfer the records via USB or Bluetooth to the SQCX quality management software for archiving.
Troubleshooting Deviations and Errors
Maintenance & Care for Maximum Lifespan
Regular care secures the measurement precision of the YS6002 over years of operation. Observe these core rules:
- Dust Protection: Always cover the measurement chamber when not in use to prevent dust accumulation inside the 154 mm integrating sphere.
- Tile Preservation: Store calibration standards in a dark, dry, and dust-protected environment inside their original case. Never touch the active measuring surface.
- Climate Stability: Operate the benchtop unit in a climate-controlled laboratory environment away from direct sunlight, drafts, and radiators.
- Gentle Touchscreen Use: Use only clean fingertips or a dedicated stylus on the capacitive 7-inch TFT display – never sharp laboratory tools.
- Recertification: Send the instrument for factory recalibration at regular intervals according to your ISO quality standards to ensure data traceability.
Frequently Asked Questions from Lab Practice (FAQ)
Question 1: Can the YS6002 measure both liquid and solid samples?
Answer: Yes. Thanks to its advanced d/0 transmissive geometry, solid materials like plastic plates and films, as well as liquids in standard cuvettes, can be precisely analyzed.
Question 2: What is the difference between the Pt-Co and Gardner indices?
Answer: The Pt-Co index (Apha value) is used for nearly clear, slightly yellowish liquids (e.g., water, solvents). The Gardner index covers darker yellow and brown tones (e.g., oils, resins). The YS6002 supports both indices natively.
Question 3: How often should calibration be performed?
Answer: As a general rule: At least once per shift or every 4 to 8 hours. If the laboratory experiences significant temperature changes, calibration should be repeated immediately.
Question 4: Does the device support measurement according to ASTM D1003?
Answer: Yes, the YS6002 fully complies with ASTM D1003/1044 requirements for evaluating haze and total transmission.
Question 5: Is the software included in the scope of delivery?
Answer: Yes, the professional quality management software SQCX is available for download, enabling extensive data analysis and report generation on a PC.
Conclusion
The 3nh YS6002 benchtop color and haze meter sets benchmarks for accuracy and user comfort. By combining grating spectroscopy with a standard-compliant d/0 geometry, laboratories eliminate uncertainties when evaluating transparent or opaque products. By maintaining the optimized workflow and regular maintenance cycles, the system delivers uncompromised, high-precision measurement results for the most demanding quality standards over years of service.